DOTX-45 Dissolved Oxygen System Part 6 Configuration
DOTX-45 45 -
When failsafe is selected to be ON, the normally-open
contacts of the relay will be closed during normal operation.
In an attempt to make this configuration less confusing, the
LCD icon logic is reversed with this setting, and the icon is
OFF under this normal condition. Therefore, when the trip
condition occurs, the closed N.O. contacts will be opened
(relay de-energized), and the LCD icon will illuminate. In
addition, a power fail would also cause the same contacts to
open.
Set Default The Set Default function allows the user to return the
instrument back to factory default data for all user settings or
for just the calibration default. It is intended to be used as a
last resort troubleshooting procedure. All user settings or
the calibration settings are returned to the original factory
values. Hidden factory calibration data remains unchanged.
Press ENTER to initiate user entry mode and select either
CAL or ALL with the UP arrow key. The default CAL routine
will reset the zero offset to 0.0 nA and reset the slope to
100%. The default ALL routine will reset all program
variables to factory default and should be used with care
since it will change any user settings that were programmed
in the field.
.