DOTX-45 Dissolved Oxygen System Part 5 Sensor Assembly
DOTX-45 21 -
Two sets of SPDT relay contacts are provided on the power supply board. None
of the relay contacts are powered. The user must supply the proper power to the
contacts. For applications that require the same switched operating voltage as
the DOTX-45, power may be jumpered from the power input terminals at TB5.
Relay wiring is connected at TB3 as shown below. Note that the relay contact
markings are shown in the NORMAL mode. Programming a relay for Failsafe
operation reverses the NO and NC positions in this diagram.
Figure 4-4 Relay Contacts
4.4 Sensor Wiring
The sensor cable can be quickly connected to the DOTX-45 terminal strip by
matching the wire colors on the cable to the color designations on the label in the
monitor. A junction box is also available to provide a break point for long sensor
cable runs. Route signal cable away from AC power lines, adjustable frequency
drives, motors, or other noisy electrical signal lines. Do not run sensor or signal
cables in conduit that contains AC power lines or motor leads.
4.5 Direct Sensor Connection
Sensor connections are made in accordance with Figure 4-1. The sensor cable
can be routed into the enclosure through one of cord-grips supplied with the unit.
Routing sensor wiring through conduit is only recommended if a junction box is to
be used. Some loose cable is needed near the installation point so that the
sensor can be inserted and removed easily from the flowcell.