DOTX-45 Dissolved Oxygen System Part 2– Analyzer Mounting
DOTX-45 14 -
2.4 Panel Mount, AC Powered Monitor
Panel mounting of an AC powered monitor uses the panel mounting flange
molded into the rear section of the enclosure. Figure 2-6 provides dimensions for
the panel cutout required for mounting.
The panel mounting bracket kit must be ordered separately. This kit contains a
metal retainer bracket that attaches to the rear of the enclosure, 4 screws for
attachment of this bracket, and a sealing gasket to insure that the panel mounted
monitor provides a water tight seal when mounted to a panel.
The sealing gasket must first be attached to the enclosure. The gasket contains
an adhesive on one side so that it remains in place on the enclosure. Remove
the protective paper from the adhesive side of the gasket and slide the gasket
over the back of the enclosure so that the adhesive side lines up with the back of
the enclosure flange. Once in place, you can proceed to mount the monitor in
the panel.
Figure 2-6 115/230 VAC Panel Mount and Cut-out