DOTX-45 Dissolved Oxygen System Part 2– Analyzer Mounting
DOTX-45 12 -
2.2 Wall or Pipe Mount
A PVC mounting bracket with attachment screws is supplied with each
transmitter (see Figure 2-3 for dimensions). The multi-purpose bracket is
attached to the rear of the enclosure using the four flat head screws. The
instrument is then attached to the wall using the four outer mounting holes in the
bracket. These holes are slotted to accommodate two sizes of u-bolt that may be
used to pipe mount the unit. Slots will accommodate u-bolts designed for 1½ “or
2” pipe. The actual center to center dimensions for the u-bolts are shown in the
drawing. Note that these slots are for u-bolts with ¼-20 threads.
Figure 2-3 Wall or Pipe mount Bracket